Spectrometer

OceanOptics* HR4000

Detector Toshiba TCD1304AP linear CCD array
Scanning range: 200nm - 1100 nm
Number of Pixels 3648
Pixel Size 8 μm x 200 μm
Pixel Well Depth ~100.000  electrons
Optical Design f/4, Symmetrical crossed Czerny-Turner
Focal length 101.6 mm input and output
Aperture standard 50 µm wide slit, other sizes are possible (5, 10, 25, 50, 100, 200)
Grating Grating #HC1, 200nm - 1100nm, best efficient 200nm - 1050nm
Detector Len L4, to increases light-collection efficiency
OFLV-200-1100 filter OFLV Variable Longpass Order-sorting Filter, to block second- and third-order light
UV enhanced window UV4
Optical Fiber connector SMA 905 to 0.22 NA single-strand optical fiber
Optical resolution For 50μm slit ~1.8nm (FWHM), for different slit size 0.02nm - 8.4nm
Signal-to-noise ratio 300:1
A/D resolution 14 bit
Dark noise 12 RMS counts
Acquisition time 3.8 milliseconds to 10 seconds
Corrected linearity >99.8%
   
  Light Source
Lamp type Combined Deuterium - Tungsten Halogen Light Source (controlled indipendently)
Wavelength range 200-410 nm (deuterium), 360-2000 nm (tungsten halogen)
Shutter Manual or Software control
Output 3.8 watts (deuterium); 1.2 watts (tungsten halogen)
Lifetime ~1500 hours (deuterium); 1500 hours (tungsten halogen)
Time to Stabilize ~10min
   
  General characteristics
Sample size 8mm - 200mm (irregular shape)
Computer requirements PC with Windows XP/Vista/7 and a USB port available
Power requirements 115VAC/230VAC (0.6A/0.3A)
Sort cut circuit Slow Blow Fuse 5A.
Dimensions (W x L x H) Standard Size: 360mm x 400mm x 180mm, Other sizes are possible to meet particular needs
Weight ~10.5Kgr, depends of the configuration
   
   