_FR-Basic D VIS/NIR |
| |
Spectrometer 1 (VIS) |
OceanOptics* USB4000 |
| Detector |
Toshiba TCD1304AP linear CCD array |
| Scanning range: |
350nm - 1100 nm |
| Number of Pixels |
3648 |
| Pixel Size |
8 μm x 200 μm |
| Pixel Well Depth |
~100.000 electrons |
| Optical Design |
f/4, Symmetrical crossed Czerny-Turner |
| Focal length |
42mm input |
| Aperture |
standard 50 µm wide slit, other sizes are possible (5, 10, 25, 50, 100, 200) |
| Grating |
Grating #3, 350nm - 1000nm, best efficient (>30%) 350nm - 850nm |
| Detector Len |
L4, to increases light-collection efficiency |
| DET4-350-1000 Filter |
OFLV Variable Longpass Order-sorting Filter, to block second- and third-order light |
| Optical Fiber connector |
SMA 905 to 0.22 NA single-strand optical fiber |
| Optical resolution |
For 50μm slit ~2.0nm (FWHM), for different slit size 0.3nm - 10nm |
| Signal-to-noise ratio |
300:1 |
| A/D resolution |
16 bit |
| Dark noise |
50 RMS counts |
| Acquisition time |
10 milliseconds to 10 seconds |
| Corrected linearity |
>99.8% |
| |
|
|
Spectrometer 2 (NIR) |
OceanOptics* NIRQuest 512 |
| Detector |
Hamamatsu G9204-512 InGaAs linear array (cooled) |
| Scanning range: |
850nm - 1700 nm |
| Number of Pixels |
512 |
| Pixel Size |
25 µm x 500 µm |
| Pixel Well Depth |
~100.000 electrons |
| Optical Design |
f/4, Symmetrical crossed Czerny-Turner |
| Focal length |
42mm input |
| Aperture |
standard 50 µm wide slit, other sizes are possible (10, 25, 50, 100, 200) |
| Grating |
Grating NIR3, 150 l/mm, 900-1700 nm |
| Filter |
OF1-RG830 longpass NIR filter, to block second- and third-order light |
| Optical Fiber connector |
SMA 905 to 0.22 NA single-strand optical fiber |
| Optical resolution |
With 50μm slit ~3.6nm (FWHM), with different slit size 2.0nm - 12.3nm |
| Signal-to-noise ratio |
15000:1 @ 100msec integration, 13000:1 @ 1000msec |
| A/D resolution |
16 bit |
| Dark noise |
6 RMS counts @ 100msec integration, 12 RMS counts @ 1000msec |
| Acquisition time |
1 milliseconds to 120 seconds |
| Corrected linearity |
>99.8% |
| |
|
|
| |
Light Source |
| Lamp type |
Tungsten Halogen Light Source |
| Wavelength range |
360-2000 nm |
| Shutter |
Manual or software control |
| Output |
20 watts (tungsten halogen) |
| Color Temperature |
3000oK |
| Lifetime |
~2000 hours (tungsten halogen) |
| Time to Stabilize |
~10min |
| |
|
|
| |
General characteristics |
| Sample size |
8mm - 200mm (irregular shape) |
| Computer requirements |
PC with Windows XP/Vista and a USB port available |
| Power requirements |
115VAC/230VAC (0.6A/0.3A) |
| Sort cut circuit |
Slow Blow Fuse 5A. |
| Dimensions (W x L x H) |
Standard Size: 440mm x 360mm x 180mm, Other sizes are possible to meet particular needs |
| Weight |
~16Kgr, depending on the configuration |
| |
|
|
|
| |
|
|